Calibration of Reactor

1.Objective :
To lay down the procedure for the calibration of reactor used in API manufacturing.
2. Scope :
Applicable in API manufacturing industry.
3. Responsibility :
Production Chemist, Production In charge, Quality Assurance
4. Definition :
- Calibration: It is a demonstration that, a particular Instrument or equipment produces results within specified limits by comparisons with those produced by a reference or traceable standard over an appropriate range of measurements.
- Working Volume: 80 % of the equipment volume is considered as Working Volume of reactor OR volume between dead & overflow volume.
- Dead Volume: Minimum volume of reactor, which is not stirrarble.
5. Procedure :
- Basic Requirements:
-Ensure that reactor is empty & clean.
-Ensure that Bottom valve of the reactor is closed.
-Make the arrangements for calibrated water flow meter, calibrated dip rod & two hosepipes.
-Connect the two end of water flow meter with the hosepipes.
-Now, put the end of one hosepipe into the reactor & the other end of second hosepipe should be connected to the tap for water flow.
-Water flow meter should be properly clamped to note down the readings.
a. Dead Volume:
- Insert the hosepipe into the reactor & calibrated dip rod at the bottom valve of reactor.
- Take the initial reading by marking on dip rod for the identification.
- Now, start water flow & allow water to fill the reactor until it touches the leaf of the agitator.
- Consider this as second reading.
- Get difference between these two readings & find out the volume of water.
- This volume of water is considered as Dead Volume of the reactor.
- Put mark on dip rod as dead volume.
b. Working Volume :
- 80 % of the equipment volume is considered as Working Volume of reactor.
- First of all, get the Liter per centimeter (LPCM) on dip rod by identifying the amount of water required to reach the one centimeter on dip rod.
- Now, fill the reactor with water until water level gets up to the neck of the reactor (almost 80 %).
- Take the difference between this reading with dead volume.
- This will give the working volume of the reactor.
c. Minimum Stirring Volume :
- Insert the hosepipe into the reactor & calibrated dip rod at the bottom valve of reactor.
- Take the initial reading by marking on dip rod for the identification.
- Now, start water flow & allow water to fill the reactor until it touches the bottom of the agitator so that it can be dip into water.
- Consider this as second reading.
- Get difference between these two readings & find out the volume of water.
- This volume of water is considered as Minimum Stirring Volume of the reactor.
d. Minimum Sensing Volume Without Stirring:
- Ensure that agitator must not be moving or rotating.
- Fill the reactor with water until it touches to the bottom of temperature sensor.
- Mark this on dip rod.
- Get the difference between this & dead volume.
- This will give Minimum sensing volume without stirring.
e. Minimum Sensing Volume With Stirring:
- Ensure that agitator is in moving condition with its minimum speed.
- Now, start the flow of water in reactor while agitator is still moving.
- Stop the water flow once the bottom of sensor touches to the water level.
- Mark this reading.
- Now, difference of this volume with dead volume will give Minimum Sensing Volume with Stirring.
For any feedback or suggestions, kindly write to info@pharmastate.com
Dead volume is not to be deducted while calculating working volume and sensing volumes.