Calibration of Reactor

1.Objective :

 To lay down the procedure for the calibration of reactor used in API manufacturing.

2. Scope :

Applicable in API manufacturing industry.

3. Responsibility :

 Production Chemist, Production In charge, Quality Assurance

4. Definition :

  • Calibration: It is a demonstration that, a particular Instrument or equipment produces results within specified limits by comparisons with those produced by a reference or traceable standard over an appropriate range of measurements.
  • Working Volume: 80 % of the equipment volume is considered as Working Volume of reactor OR volume between dead & overflow volume.
  • Dead Volume: Minimum volume of reactor, which is not stirrarble.

5. Procedure :

  • Basic Requirements:

-Ensure that reactor is empty & clean.

-Ensure that Bottom valve of the reactor is closed.

-Make the arrangements for calibrated water flow meter, calibrated dip rod & two hosepipes.

-Connect the two end of water flow meter with the hosepipes.

-Now, put the end of one hosepipe into the reactor & the other end of second hosepipe should be connected to the tap for water flow.

-Water flow meter should be properly clamped to note down the readings.

a. Dead  Volume:

  • Insert the hosepipe into the reactor & calibrated dip rod at the bottom valve of reactor.
  • Take the initial reading by marking on dip rod for the identification.
  • Now, start water flow & allow water to fill the reactor until it touches the leaf of the agitator.
  • Consider this as second reading.
  • Get difference between these two readings & find out the volume of water.
  • This volume of water is considered as Dead Volume of the reactor.
  • Put mark on dip rod as dead volume.

b. Working Volume :

  • 80 % of the equipment volume is considered as Working Volume of reactor.
  • First of all, get the Liter per centimeter (LPCM) on dip rod by identifying the amount of water required to reach the one centimeter on dip rod.
  • Now, fill the reactor with water until water level gets up to the neck of the reactor (almost 80 %).
  • Take the difference between this reading with dead volume.
  • This will give the working volume of the reactor.

c. Minimum Stirring Volume :

  • Insert the hosepipe into the reactor & calibrated dip rod at the bottom valve of reactor.
  • Take the initial reading by marking on dip rod for the identification.
  • Now, start water flow & allow water to fill the reactor until it touches the bottom of the agitator so that it can be dip into water.
  • Consider this as second reading.
  • Get difference between these two readings & find out the volume of water.
  • This volume of water is considered as Minimum Stirring Volume of the reactor.

d. Minimum Sensing Volume Without Stirring:

  • Ensure that agitator must not be moving or rotating.
  • Fill the reactor with water until it touches to the bottom of temperature sensor.
  • Mark this on dip rod.
  • Get the difference between this & dead volume.
  • This will give Minimum sensing volume without stirring.

e. Minimum Sensing Volume With Stirring:

  • Ensure that agitator is in moving condition with its minimum speed.
  • Now, start the flow of water in reactor while agitator is still moving.
  • Stop the water flow once the bottom of sensor touches to the water level.
  • Mark this reading.
  • Now, difference of this volume with dead volume will give Minimum Sensing Volume with Stirring.

For any feedback or suggestions, kindly write to info@pharmastate.com

One Comment

  • PharmaState Academy
    17/04/2019

    Dead volume is not to be deducted while calculating working volume and sensing volumes.

Write a comment